AC/DC Medical Desktop
130 Watts
PMP130 Series
XPiQ inc.
Intelligent Design Quality Product
•
CE Marked - LVD & EMC
•
Universal Input
•
100% Burn-In
•
Low Leakage Current
•
4000 VAC Isolation
•
Input Surge Protection
•
Line & Neutral Fused
Specification
All specifications typical at nominal line, full load and 25°C
Input
Input Voltage
Input Frequency
Input Current
Safety Ground
Leakage Current
• 85-130 VAC or 180 to 264 VAC
auto ranging
• 47 to 63 Hz
• 3.0 A (rms) for 115 VAC
1.7 A (rms) for 230 VAC
• 50 µA max at 110 VAC 60 Hz
• 100 µA max at 240 VAC 50 Hz
Output
Total Output Power •
Ripple and Noise
•
Overvoltage
•
Protection
Overcurrent
•
Protection:
Temperature
•
Coefficient:
Transient Response •
Remote Sense
•
See Tables
1% pk-pk max
Set at 112-132% of its nominal
output voltage
Output protected to short
circuit conditions
±0.04%/°C
Maximum excursion of 4% or
better on all models. Recovers
to 1% of final value within
500 µs after a 25% step load
change
Optional, please consult factory
General
Efficiency
•
Hold-up time
•
Line Regulation
•
Inrush Current
•
Withstand Voltage •
Insulation Resistance •
Switching Frequency •
Environmental
Operating
•
Temperature
Storage
•
Temperature
Relative Humidity •
MTBF
•
EMC & Safety
EMI Requirements •
Safety Requirements •
70% minimum
10 msec minimum at 110 VAC
±0.5% maximum at full load
15 A at 115 VAC or
30 A at 230 VAC at +25°C cold start
4000 VAC from input to output
1500 VAC from input to ground
500 VAC from output to ground
10 Mohm minimum from
output to ground
60 kHz to 75 kHz
0 °C to +70 °C, derate linearly from
100% load at +50 °C to 50% load
at +70 °C
-40 °C to +85 °C
5% to 95% non-condensing
100,000 hours minimum at full load
at +25 °C ambient
In compliance with radiated and
conducted limits of CISPR 11
(EN 55011) Class B
Approved to: (not for patient contact)
a) UL 2601
b) CSA C22.2 No. 601.1
c) IEC 601-1/EN60601-1
•
www.XPiQ.com
PH: 508 429.9883 •
FAX: 800 226.2100
Email: sales@xpiq.com
Holliston, MA 01746 USA