Data Sheet
Parameter
Min
Gain Error
−1
−2
Gain Matching
−0.1
−0.5
Gain Tempco1
Input Referred Noise1
Analog Lead Mode
Electrode Mode
Digital Lead Mode
Power Supply Sensitivity2
Analog Channel Bandwidth1
Dynamic Range1
Signal-to-Noise Ratio1
COMMON-MODE INPUT
Input Voltage Range
Input Impedance2
Input Bias Current
COMMON-MODE OUTPUT
VCM_REF
Output Voltage, VCM
Output Impedance1
Short Circuit Current1
Electrode Summation
Weighting Error2
RESPIRATION FUNCTION
(ADAS1000 ONLY)
Input Voltage Range
Input Voltage Range (Linear
Operation)
Input Bias Current
Input Referred Noise1
Frequency2
Excitation Current
0.3
−40
−200
1.28
0.3
0.3
−10
Resolution2
Measurement Resolution1
In-Amp Gain1
Gain Error
Gain Tempco1
Typ
+0.01
+0.1
+0.02
+0.1
25
6
10
12
11
12
14
16
100
65
104
100
1||10
±1
1.3
1.3
0.75
4
1
Max
+1
+2
+0.1
+0.5
2.3
+40
+200
1.32
2.3
2.3
1.8/gain
±1
+10
0.85
46.5 to 64
64
32
16
8
24
0.2
0.02
1 to 10
1
25
ADAS1000/ADAS1000-1/ADAS1000-2
Unit
%
%
%
%
ppm/°C
μV p-p
μV p-p
μV p-p
μV p-p
μV p-p
μV p-p
μV p-p
dB
kHz
dB
dB
V
GΩ||pF
nA
nA
V
V
mA
%
V
V p-p
nA
μV rms
kHz
μA p-p
μA p-p
μA p-p
μA p-p
bits
Ω
Ω
%
ppm/C
Test Conditions/Comments
GAIN 0 to GAIN 2, factory calibrated. Programmable
user or factory calibration option enables. Factory gain
calibration applies only to standard ECG interface.
GAIN 3 setting, no factory calibration for this gain
GAIN 0 to GAIN 2
GAIN 3
GAIN 2, 2 kHz data rate, see Table 4
0.5 Hz to 40 Hz; high performance mode
0.05 Hz to 150 Hz; high performance mode
0.05 Hz to 150 Hz; low power mode
0.05 Hz to 150 Hz; high performance mode
0.05 Hz to 150 Hz; low power mode
0.05 Hz to 150 Hz; high performance mode
0.05 Hz to 150 Hz; low power mode
At 120 Hz
GAIN 0, 2 kHz data rate, −0.5 dBFS input signal, 10 Hz
−0.5 dB FS input signal
CM_IN pin
Over operating range; dc and ac lead-off disabled
AGND to AVDD
CM_OUT pin
Internal voltage; independent of supply
No dc load
Not intended to drive current
Resistor matching error
These specifications apply to the following pins:
EXT_RESP_LA, EXT_RESP_LL, EXT_RESP_RA and selected
internal respiration paths (Lead I, Lead II, Lead III)
AC-coupled, independent of supply
Programmable gain (10 states)
Applies to EXT_RESP_xx pins over AGND to AVDD
Programmable frequency, see Table 28
Respiration drive current corresponding to differential
voltage programmed by RESPAMP bits in RESPCTL
register. Internal respiration mode, cable 5 kΩ/200 pF,
1.2 kΩ chest impedance
Drive Range A
Drive Range B2
Drive Range C2
Drive Range D2
Update rate 125 Hz
Cable <5 kΩ/200 pF per electrode, body resistance
modeled as 1.2 kΩ
No cable impedance, body resistance modeled as 1.2 kΩ
Digitally programmable in steps of 1
LSB weight for GAIN 0 setting
Rev. A | Page 5 of 80