5
High Speed Testing of Chip Inductors and
Magnetic Heads in Research and Development
3535 Specifications
Measurement
Items
: Z (impedance), Y (admittance), Rs (series-
equivalent resistance, ESR), Rp (parallel-
equivalent resistance), G (conductivity), X
(reactance), B (susceptance), θ (phase angle),
Ls (series-equivalent inductance), Lp (parallel-
equivalent inductance), Cs (series-equivalent
capacitance), Cp (parallel-equivalent capacitance),
Q (Q factor), D (loss constant tanδ)
Measurement
Frequency
: Frequency Range: 100 kHz to 120 MHz
Setting Resolution: Four digits (by front panel setting)*
100.0 kHz to 1.000 MHz: 100-Hz steps
1.000 MHz to 10.00 MHz: 1-kHz steps
10.00 MHz to 100.0 MHz: 10-kHz steps
100.0 MHz to 120.0 MHz: 100-kHz steps
*1-Hz resolution with GP-IB or RS-232C interface
Frequency Accuracy: less than ±0.005% of
setting value
Output Impedance : 50 ±10 Ω (at 100 kHz)
Measurement
Signal Level
: Open-terminal voltage (V) mode
Level Range:
5 mV to 1 V, 20 mA max. (up to 10.00 MHz)
5 mV to 500 mV, 10 mA max. (above 10.01 MHz)
Setting Resolution: 1mV steps
Setting Accuracy: ±(5% + 5 mV) × (2 + log f)
where f is in MHz
Constant Current (CC) mode
Level Range:
200 µA to 20 mA: 1 V max. (up to 10.00 MHz)
200 µA to 10 mA: 0.5 V max. (above 10.01 MHz)
Setting Resolution: 10 µA steps
Setting Accuracy: ±(10% + 50 µA) × (2 + log f)
where f is in MHz
Monitor Function : Monitor Voltage: 0.000 to 1.000 V
Monitor Current: 0.000 to 20.00 mA
Limit Function : Current Limit (during V setting): 0.20 to 20.00 mA
Voltage Limit (during CC setting): 0.005 to 1.000 V
Measurement
Time
: 6 ± 1 ms (nominal)
Actual time depends on measurement conditions,
such as measurement speed and averaging.
Measurement
Speed
: FAST, NORMAL, SLOW and SLOW2
Average
: OFF, 2, 4, 8, 16, 32 and 64
Trigger Function : Internal and external trigger sources can be selected.
Trigger Delay function:
0.01 to 9.99 s with 0.01 s resolution
Load Compensation : Measure a standard component to establish
Function
a compensation value for subsequent
measurements.
Key Lock
Function
: Temporarily disable touch panel operation
using rear panel switch.
Comparator
: Compares two measurement items.
Input either the absolute value, percentage of
standard value or its deviation percentage (Δ%).
Note: for Δ%, the measurement value is
displayed as the percentage of deviation
from the standard value.
Classification (BIN) : Ranks two measurement items into ten
Measurement
classifications.
Correlation
: Constants a and b are entered to compensate
Compensation Function displayed values.
[Compensated value] = a × [measurement value] + b
Panel Save and
Load
: Memory Capacity: 30 Sets
Load Method: Front panel key operation,
External I/O connector, GP-IB, RS-232C
Measurement Value : Memory Capacity: Up to 200 values
Storage
Measurement values are stored in the main unit
and transferred as a batch.
Zoom Display
Function
: Enlarge the display of measurement data and
comparator results.
Continuous
Measurements
: Measurements are made continuously per
conditions saved on the screen.
Audible Beeper
: Beeping can be set ON/OFF for key entry and
comparator results (IN or NG).
Numerical Display
Digit Setting
Function
: Measurement values can be set to display as 3,
4 or 5 digits.
Available settings depend on the parameter.
Display Setting
Functions
: Backlight and voltage/current monitor display
can be set ON/OFF.
Note: when the backlight is off, display refresh
is disabled (during high-speed measurement).
Printer Functions
: Hard copy printout of measurement values and
screens.
Note: requires 9442 and 9444
Interfaces
: GP-IB, RS-232C and EXT I/O (standard)
Operating
Temperature and
Humidity
: 10 to 40°C, 80% rh or less, no condensation
Storage
Temperature and
Humidity
: -10 to 55°C, 80% rh or less, no condensation
Operating
Environment
: Indoors, up to 2000 m ASL
Power
: 100 to 240 VAC, 50/60 Hz
Maximum Rated : 50 VA
Power
Dimensions and Mass : Approx. 360W × 130H × 360D mm, 8.3 kg
Conforming
Standards
: EMC: EN61326:1997+A1:1998+A2:2001
EN61000-3-2:2000
EN61000-3-3:1995+A1:2001
Safety: EN61010-1:2001
Pollution Degree 2