CS5531/32/33/34-AS
TYPICAL RMS NOISE (nV), CS5531/32/33/34
(See notes 11, 12 and 13)
Output Word
Rate (Sps)
7.5
15
30
60
120
240
480
960
1,920
3,840
-3 dB Filter
Frequency (Hz)
1.94
3.88
7.75
15.5
31
62
122
230
390
780
x64
17
24
34
48
68
115
163
229
344
1390
x32
17
25
35
49
70
160
230
321
523
2710
Instrumentation Amplifier Gain
x16
x8
x4
19
26
42
27
36
59
39
51
84
54
72
118
77
102
167
276
527
1040
392
748
1480
554
1060
2090
946
1840
3650
5390 10800 21500
x2
79
111
157
222
314
2070
2950
4170
7290
43000
x1
155
218
308
436
616
4150
5890
8340
14600
86100
Notes: 11. Wideband noise aliased into the baseband. Referred to the input. Typical values shown for 25 °C.
12. For peak-to-peak noise multiply by 6.6 for all ranges and output rates.
13. Word rates and -3dB points with FRS = 0. When FRS = 1, word rates and -3dB points scale by 5/6.
TYPICAL NOISE-FREE RESOLUTION(BITS), CS5532/34 (See Notes 14 and 15)
Output Word -3 dB Filter
Instrumentation Amplifier Gain
Rate (Sps) Frequency (Hz) x64
x32
x16
x8
x4
x2
x1
7.5
1.94
19
20
21
22
22
22
22
15
3.88
19
20
21
21
21
22
22
30
7.75
18
19
20
21
21
21
21
60
15.5
18
19
20
20
20
21
21
120
31
17
18
19
20
20
20
20
240
62
16
17
17
17
17
17
17
480
122
16
17
17
17
17
17
17
960
230
15
16
16
16
16
16
16
1,920
390
15
15
15
15
15
15
15
3,840
780
13
13
13
13
13
13
13
14. Noise-free resolution listed is for bipolar operation, and is calculated as LOG((Input Span)/(6.6xRMS
Noise))/LOG(2) rounded to the nearest bit. For unipolar operation, the input span is 1/2 as large, so one
bit is lost. The input span is calculated in the analog input span section of the data sheet. The noise-free
resolution table is computed with a value of 1.0 in the gain register. Values other than 1.0 will scale the
noise, and change the noise-free resolution accordingly.
15. “Noise-free resolution” is not the same as “effective resolution”. Effective resolution is based on the
RMS noise value, while noise-free resolution is based on a peak-to-peak noise value specified as 6.6
times the RMS noise value. Effective resolution is calculated as LOG((Input Span)/(RMS
Noise))/LOG(2).