Definitions
Specification (spec): Represents warranted performance of a calibrated instrument
that has been stored for a minimum of 2 hours within the operating temperature range of
0 to 55 °C, unless otherwise stated, and after a 45 minute warm-up period. The specifi-
cations include measurement uncertainty. Data represented in this document are
specifications unless otherwise noted.
Typical (typ): Represents characteristic performance, which 80% of the instruments
manufactured will meet. This data is not warranted, does not include measurement
uncertainty, and is valid only at room temperature (approximately 25 °C).
Nominal (nom): The expected mean or average performance, or an attribute whose
performance is by design, such as the 50 Ω connector. This data is not warranted and is
measured at room temperature (approximately 25 °C).
Measured (meas): An attribute measured during the design phase for purposes of
communicating expected performance, such as amplitude drift vs. time. This data is not
warranted and is measured at room temperature (approximately 25 °C).
Note: All graphs contain measured data from several units at room temperature unless
otherwise noted.