datasheetbank_Logo
búsqueda de Hoja de datos y gratuito Fichas de descarga

N5162A Ver la hoja de datos (PDF) - HP => Agilent Technologies

Número de pieza
componentes Descripción
Lista de partido
N5162A
HP
HP => Agilent Technologies HP
N5162A Datasheet PDF : 32 Pages
1 2 3 4 5 6 7 8 9 10 Next Last
Frequency
Range
Option 503
Option 506
100 kHz to 3 GHz
100 kHz to 6 GHz
Minimum frequency
100 kHz 1
Resolution
0.01 Hz
Phase offset
Adjustable in nominal 0.01 ° increments
Frequency bands 2
Band
1
2
3
4
5
6
Frequency range
N
100 kHz to < 250 MHz
1
250 to < 375 MHz
0.25
375 to < 750 MHz
0.5
750 to < 1500 MHz
1
1500 to < 3000.001 MHz
2
3000.001 to 6000 MHz
4
Switching speed 3, 4
Type
Standard
Digital modulation off
SCPI mode
List/Step sweep mode
Digital modulation on
SCPI mode
List/Step sweep mode
≤ 5 ms (typ)
≤ 5 ms (typ)
≤ 5 ms (typ)
≤ 5 ms (typ)
Option UNZ 5
≤ 1.15 ms
≤ 900 µs
≤ 1.15 ms
≤ 900 µs
Option UNZ 5
(typical)
≤ 950 µs
≤ 700 µs
≤ 1.05 ms
≤ 800 µs
1. Performance below 250 kHz is unspecified except as indicated, for units with serial numbers
ending with 4742xxxx or greater. For units with lower serial numbers refer to the Archive
Section at end of this document.
2. N is a factor used to help define certain specifications within the document.
3. Time from receipt of SCPI command or trigger signal to within 0.1 ppm of final frequency
or within 100 Hz, whichever is greater, and amplitude settled to within 0.2 dB.
4. Additional time may be required for the amplitude to settle within 0.2 dB when switching
to or from frequencies < 500 kHz.
5. Specifications apply when status register updates are off.
4